Technical Notes:

• Horizontal surface mount (±30°) recommended for

optimum performance.

• Suitable for clean fl uids only. Suspended solids will cause

mechanical failure.

• 1/4 in. NPT or ISO 7/1 - R1/4 (male) port connections

• NEMA 4X/IP65 splashproof enclosure

2. Recommended Sensor Position

1. Compatible Instrument Wiring

Technical Notes:
•

Use 2-conductor twisted-pair shielded cable for sensor cable splices up to 300 m (1000 ft) max.

•

Maintain cable shield through cable splice.

•

Route sensor cable away from AC power lines.

•

AUX power required when used with any Signet 8550 Process Pro Flow Transmitter.

Signet 2000

Input

Gnd.

10 k

Ω

+

5 to 24
VDC

-

Black

Shield

Red

Other
Instrument

Black (5 to 24 VDC)

Red (Signal out)

Shield (DC return)

Blk
Power

Red
Freq. input
Shld
Gnd

Other

Instrument

Signet

Instrument

0°

30°

30°

Recommended

mounting angles

Vertical mounting

not recommended

SAFETY INSTRUCTIONS
1. Depressurize and vent system prior to installation/removal.
2. Confi rm chemical compatibility before use.
3. Do not exceed maximum temperature/pressure specifi cations.
4. Wear safety goggles or faceshield during installation/service.
5. Do not alter product construction.