Warranty, Specifications, Temperature channel – Green Air Products Telaire 7001 User Manual

Page 2: Power supply, Telaire, 7001 carbon dioxide and temperature monitor, Performance

Advertising
background image

Warranty

This Engelhard product has been examined
and tested for proper operation. Please op-
erate this product only in accordance with
the instructions.

Engelhard warrants this product against de-
fects in workmanship and materials for a pe-
riod of 18 months from the date of purchase
by the original owner. If the product should
become defective within this warranty period
we will repair or exchange it. A return autho-
rization number must be obtained from the
factory prior to returning equipment. Items re-
ceived without a return authorization number
will be refused. Product to be serviced under
this warranty should be sent to Engelhard,
6489 Calle Real, Goleta, CA 93117. Ship-
ment must be prepaid, properly packed and
insured.

Engelhard is not and will not be liable for
any consequential loss or damages that
may occur by reason of purchase and use
of this product. The responsibility of Engel-

hard, in any event, is strictly limited to the
replacement/repair of the product.

Engelhard seeks to present reliable informa-
tion concerning the composition, properties
and use of its products, however; (1) All ad-
vice concerning selection and use of any
product is provided at no charge and with no
warranty. (2) No warranty is made hereby.
Products described herein are warranted to
conform to Engelhard’s specifications only at
the time of sale. All sales are subject to
Engelhard’s standard terms and conditions,
which are reproduced on the reverse side of
each invoice. All warranties of merchantabil-
ity and fitness of purpose are disclaimed and
remedy for any breach of warranty is limited
to replacement of the defective product. (3)
Engelhard assumes no responsibility for any
patent liability arising from the use of any
product in a process, manner or formula not
designed by Engelhard.

Specifications

Method

Dual Beam Absorption Infrared™

Sample Method

Diffusion or flow through (50 - 100 ml/min)

Warranty

18 months parts and labor

Performance

CO2 Channel
Measurement Range

0-4,000 ppm voltage output
0-10,000 ppm display

Sensitivity

± 1 ppm

Accuracy

±50 ppm or ±5% of reading, whichever is
greater

Repeatability

±20 ppm

Temperature Dependence

±0.1% of reading per °C or ±2 ppm per °C,
whichever is greater, referenced to 25°C

Pressure Dependence:

0.13% of reading per mm Hg
(Corrected via user input for elevation)

Annual Drift

± 20 ppm typical

Response Time

<60 seconds for 90% of step change

Warm-Up Time

<60 seconds at 22°C

Operating Conditions

32-122°F (0-50°C)
0-95% RH, non-condensing

Storage Temperatures

-40 to 140°F (-40 to 60°C)

Calibration Interval

12 months, offset adjustment using single
gas at 0-1000 ppm CO2. Full factory
calibration available

Temperature Channel

Temperature Range

Voltage output 32 to 104°F (0 to 40°C)
Display 32 to 122°F (0 to 50°C)

Display Resolution

0.1°F (0.1°C)

Display Options

°F, °C, or Off. Set with panel button.

Accuracy

±2°F (±1°C)

Response Time

20-30 minutes (case must equilibrate with
environment)

Calibration Interval

12 months, offset adjustment using
temperature standard at 50 to 86°F (10 to
30°C). Full factory calibration available

Output - Analog

CO2

0-4 VDC, 1mV/ppm (4,000 ppm max)

Temperature

0-4 VDC linear, 32-104°F (0-40°C)

Output Impedance

100 Ohms

Wiring Connection

Via RJ-45 to (accessory cable 2070 or
2071)

Output - Digital

Wiring Connection

Via RJ-45 to DB9 serial port cable

Display

LCD with independent CO2/ temperature
readings (panel buttons set elevation, °F/
°C, calibration functions)

Power Supply

Battery Type
Four AA batteries, not included
Battery Operation

48 hours (alkaline)

External

6 VDC from external AC/DC adapter,
included

Power Requirements

100 mA Peak, 20 mA average from 6V

Certification

FCC Class 15 Part B

Accessories

2070 Datalogging cable for analogoutput
2071 Datalogging cable Recordaire

®

2075 Calibration kit
1058 Recordaire

®

datalogger, 4-channel

1058GSW VG Graphing software

All specifications and instructions subject to
change without notice. Protected by United
States Patents - 5,060,508 and 5,163,332.

Telaire

®

7001

Carbon Dioxide and Temperature Monitor