5.4 D/A Calibration

In a way similar to the ADCAL.EXE program, the DACAL.EXE
program leads you through the whole D/A calibration procedure.

You can either use the on-board -5 V (-10 V) internal reference
voltage or use an external reference. If you use an external reference,
connect a reference voltage within the range ±10 V to the reference
input of the D/A output channel you want to calibrate. Adjust the full
scale (gain) of D/A channel 0 and 1, with VR4 and VR5 respectively.

Note!

Using a precision voltmeter to calibrate the D/A
outputs is recommended.

Set the D/A data register to 4095 and adjust VR3 until the D/A output
voltage equals the reference voltage minus 1 LSB, but with the
opposite sign. For example, if V

ref

is -5 V, then V

out

should be +4.9959

V. If V

ref

is -10 V, V

out

should be +9.9918 V.

A/D code

Mapping Voltage

Hex.

Dec.

Bipolar ± 5V

Unipolar 0 to 10V

000h

0

-4.9971V

0V

7FFh

2047

-0.0024V

4.9947V

800h

2048

0V

4.9971V

FFFh

4095

+4.9947V

9.9918V