Specifications

on page 3

General information

on page 4

Installation

on page 7

Specifications

Specifications are subject to change without notice.

Table 1 General specifications

Specification

Details

Dimensions (W x D x H)

452 x 360 x 804 mm (17.8 x 14.2 x 31.7 in.)

Enclosure

Rating: NEMA 4x/IP65
Material: PC/ABS case, PC door, PC hinges and
latches, 316 SST hardware
Indoor use only. Keep away from direct sunlight.

Weight

20 kg (45 lb) without reagents and standards,
36.3 kg (80 lb) with reagents

Mounting

Wall, panel or table

Protection class

I

Pollution
degree/installation
category

2/II

Power requirements

AC: 100–240 VAC, 50/60 Hz
Instrument: 0.5 A nominal, 2.6 A maximum
Connection: 18 to 16 AWG wire, 18 AWG stranded
recommended

Operating temperature

5 to 45 °C (41 to 113 °F)

Operating humidity

5 to 95% non-condensing

Storage temperature

–20 to 60 °C (–4 to 140 °F)

Table 1 General specifications (continued)

Specification

Details

Maximum altitude

2000 m (6560 ft)

4–20 mA outputs

Four; load impedance: 600 Ω maximum
Connection: 22 to 16 AWG wire, 22 to 20 AWG
recommended, twisted pair shielded wire

Alarm relay outputs

Four; type: not powered SPDT relays, each rated at
5 A resistive, 240 VAC maximum
Connection: 18 to 16 AWG wire, 18 AWG stranded
recommended

Digital inputs

Four; connection: 22 to 16 AWG wire, 22 to
20 AWG stranded (isolated DC voltage input or an
open-collector/relay contact closure input)
recommended

Fuses

Input power—AC: T 1.6 A, 250 VAC; DC: T 6.3 A,
250 VAC
Output power—AC: T 5.0 A, 250 VAC; DC: T 1.6 A,
250 VAC
Alarm relay outputs: T 5.0 A, 250 V

Fittings

Sample line and sample bypass drain: 6 mm OD
push-to-connect fitting for plastic tubing
Air purge air inlet: 6 mm OD push-to-connect fitting
for plastic tubing
Chemical and case drains: 11 mm (7/16 in.) ID slip-
on fitting for soft plastic tubing

Sample pressure, flow
rate, and temperature

Pressure: 2–87 psi to preset pressure regulator
Flow rate: 55–300 mL/minute
Temperature: 5 to 50 °C (41 to 122 °F)

Number of sample streams 1, 2 or 4; programmable sequence

Air purge (optional)

0.425 m

3

/hour (15 scfh

1

), instrument quality air

Certifications

ETL certified to UL and CSA standards, CE marked

1

standard cubic feet per hour