Set the error hold mode, View data – Hach-Lange HACH 5500 sc SiO2 Operations User Manual

Page 14

Advertising
background image

Option

Description

OffMax TIMER

Sets the maximum time the relay stays off (default =
0 min).

OnMin TIMER

Sets the time the relay stays on, independent from the
measured value (default = 0 min).

OffMin TIMER

Sets the time the relay stays off, independent from the
measured value (default = 0 min).

6. Select SCHEDULER in the FUNCTION menu and then select the

applicable options in the ACTIVATION menu.

Option

Description

HOLD OUTPUTS Holds or transfers outputs for the selected channels.

RUN DAYS

Selects the days the relay stays on. Options: Sunday,
Monday, Tuesday, Wednesday, Thursday, Friday,
Saturday

START TIME

Sets the start time.

INTERVAL

Sets the time between activation cycles (default = 5 min).

DURATION

Sets the period of time the relay is set to on (default =
30 sec).

OFF DELAY

Sets the time for additional hold/output time after the
relay is set to off.

7. Select WARNING in the FUNCTION menu and then select the

applicable options in the ACTIVATION menu.

Option

Description

WARNING LEVEL Sets the level for the warning activation and starts

applicable individual warnings.

Set the error hold mode

1. Select SETUP SYSTEM>SETUP OUTPUTS>ERROR HOLD MODE.

2. Select an option.

Option

Description

HOLD OUTPUTS

Holds the outputs at the last known value when
communications are lost.

TRANSFER
OUTPUTS

Switches to the transfer mode when communications
are lost. The outputs transfer to a pre-defined value.

View data

1. Select VIEW DATA.
2. Select an option.

Option

Description

ANALYZER DATA

Shows specific status information (refer to

Table 2

).

MEASUREMENT
DATA

Shows measurement information (refer to

Table 3

).

LOG DATA

Selects the data log and the event log. DATA LOG
—shows the measurement values. Select start
time, number of hours and number of readings.
EVENT LOG—shows all analyzer information.
Select start time, number of hours and number of
readings.

Table 2 Analyzer data

Element

Definition

CELL TEMP

Temperature of the colorimeter cell block heater (ideally
49.8 °C to 50.2 °C (121.64 °F to 122.36 °F))

REAGENT TEMP

Reagent temperature before it enters the colorimeter

AMBIENT TEMP

Air temperature inside the electronics area

SAMPLE TEMP

Sample pre-heater block temperature (typically 45 °C to
55 °C (113 °F to 131 °F) but could reach 58 °C
(136.4 °F))

14 English