Start the analyzer setup

When the analyzer is powered on for the first time, or powered on after
the configuration settings have been set to their default values:

1. Select the applicable LANGUAGE.
2. Select the DATE FORMAT.
3. Set the DATE and TIME.
4. When prompted, confirm with YES to setup the analyzer.

Note: The analyzer stays in initialization mode until the configuration is
complete.

5. Select the channel.
6. Select the measurement mode.
7. The setup results for the channel, sample flow, sample pressure and

minimum flow are shown. Make sure that the following values are
within range:

• Sample pressure: minimum 0.14 bar (2 psi)

Note: The maximum pressure is regulated by a pressure regulator at
0.28 bar (4 psi).

• Minimum flow: 55 mL/minute

8. Push enter to confirm.
9. When prompted, make sure that the reagent bottles are full and

confirm with YES.

10. When prompted, make sure that the standard bottles are full and

confirm with YES.

11. Reagent 1 has a known effect on the measurement value. The value

attributed to this effect is printed on the bottle as a BLANK value.
When prompted, enter this value.

Note: The analyzer setup does not configure relays, outputs, network cards,
calculations or calibration parameters. Refer to

Calibration

on page 17 for

configuration information for calibration parameters. Refer to

Configure the outputs

on page 11 for configuration information for relays, outputs or network cards.
Refer to

Set up the calculation

on page 11 to setup calculations.

Configure the sequencer (optional)

Option only available for the multi-channel version of the analyzer.

1. Select SETUP SYSTEM>CONFIGURE SEQUENCER.

Option

Description

ACTIVATE
CHANNELS

Starts or stops measurements for individual sample
sources. Use the UP and DOWN navigation keys to scroll
through the channels. Deselect a channel with the LEFT
navigation key. Push enter to confirm.
Note: Inactive channels are shown with a "~" character
before the channel name on all displays.

SEQUENCE
CHANNELS

Sets the measurement order of the sample sources. Use
the UP and DOWN navigation keys to scroll through the
sequence. For each sequence number use the LEFT and
RIGHT navigation keys to select a channel. Push enter to
confirm.

Calibrate the analyzer

The manufacturer recommends calibration of the analyzer after 1 day of
operation to allow all system components to stabilize. Refer to

Calibration

on page 17 to start a calibration.

Operation

W A R N I N G

Potential fire and explosion hazard. This equipment is intended for only aqueous
type samples. Use with flammable samples can result in a fire or explosion.

C A U T I O N

Chemical hazard. Always follow appropriate laboratory safety
procedures when chemicals are handled. Always wear all personal
protective equipment appropriate to the chemicals used.

When the lower door is opened, the current measurement or calibration
is stopped. When the lower door is closed, the previous measurement or
calibration starts again.