Connect to the digital inputs
The analyzer can receive a digital signal from an external device, such
as a flow meter to stop measurements when flow stops. Each discrete
input is used to enable/disable the corresponding sample channel.
Each digital input can be configured as an isolated TTL type digital input
or as a relay/open-collector type input. Refer to

Figure 15

. By default, the

jumpers are set for isolated TTL type digital input. Refer to

Wiring

connections overview

on page 13 to connect the device.

Figure 15 Isolated TTL type digital input

1 Jumper (12x)

3 Isolated TTL type digital input

2 Digital input connectors

4 Relay/Open-collector type input

Install additional modules

Modules can be added for additional output, relay or communications
options. Refer to the documentation that is supplied with the module.

English 17