LDO Sensor Replacement Kit

GB

Figure 1

Installing New Sensor Cap

8. Hold the update module to the back of the meter. Slide the rounded end of the

module until it stops at the probe holder. See Figure 2. The pins on the module
will align with the metal contact pads on the meter.

9. Sensor data is loaded automatically when the meter senses the update module.

The update takes less than two seconds.

10. After the sensor data is loaded, the meter will beep twice and display the

four-digit sensor lot code. Make sure the lot code on the meter display matches
the lot code printed on the sensor cap. The lot code begins with a number
between ”3“ and ”9“. See Figure 3.

Note: The serial number begins with a ”0“ or ”1“.

11. Press

READ/OK

to store the new sensor setup.