Calculate the pump cycle time – Hach-Lange SD900 User Manual User Manual

Page 17

Advertising
background image

W A R N I N G

Electric shock hazard. The power supply can overheat when the time between pump cycles is too
short.

N O T I C E

Make sure that the pump cycle time will not cause the power supply to overheat. Refer to

Calculate the pump

cycle time

on page 15.

Connect the sampler to a battery or to an AC power supply. Refer to the illustrated steps that follow.

Calculate the pump cycle time

The pump must stay off for a minimum time period between sample cycles so that the power supply
does not overheat.

1. Calculate the total amount of time that the pump will run continuously during a sample cycle.

Include all stages: pre-purge, intake rinse, sample, sample retries and post-purge.

2. Find the minimum amount of time that the pump must stay off between sample cycles from

Figure 11

.

3. Make sure that the pacing or time interval in the sampler program lets the pump stay off between

sample cycles for the minimum amount of time.

English 15