Specification

Details

Storage/operating humidity

100% condensing

Pump

Peristaltic high speed, with spring-mounted Nylatron rollers

Pump enclosure

IP37

Pump tubing

9.5 mm ID x 15.9 OD mm (

3

/

8

in. ID x

5

/

8

in. OD) silicone

Pump tubing life

20,000 sample cycles with: 1 L (0.3 gal) sample volume, 1 rinse, 6 minute
pacing interval, 4.9 m (16 ft) of

3

/

8

in. intake tube, 4.6 m (15 ft) of vertical lift,

21 °C (70 °F) sample temperature

Vertical sample lift

Maximum 8.5 m (28 ft) for: 8.8 m (29 ft) of

3

/

8

-in. vinyl intake tube at sea level

at 20–25 °C (68–77 °F)

Pump flow rate

4.8 L/min (1.25 gpm) at 1 m (3 ft) vertical lift with

3

/

8

-in. intake tube typical

Sample volume

Programmable in 10-mL (0.34 oz) increments from 10 to 10,000 mL (3.38 oz
to 2.6 gal)

Sample volume repeatability
(typical)

±5% of 200 mL sample volume using uncalibrated liquid detect with: 4.6 m
(15 ft) vertical lift, 4.9 m (16 ft) of

3

/

8

-in. vinyl intake tube, single bottle, full

bottle shut-off at room temperature and 1524 m (5000 ft) elevation

Sample volume accuracy
(typical)

±10% of 200 mL sample volume using uncalibrated liquid detect with: 4.6 m
(15 ft) vertical lift, 4.9 m (16 ft) of

3

/

8

-in. vinyl intake tube, single bottle, full

bottle shut-off at room temperature and 1524 m (5000 ft) elevation

Sampling modes

Pacing: Time-fixed, flow-fixed, time-variable, flow-variable, constant time
variable volume (CTVV).
Refer to

Constant Time Variable Volume (CTVV) sampling

on page 23.

Distribution: Single bottle composite, multi-bottle composite, multi-bottle
discrete, bottles per sample, samples per bottle, combination of bottles per
sample, samples per bottle

Run modes

Continuous or non-continuous with user-entered number of samples

Multiple programs

Stores up to three sampling programs

Transfer velocity (typical)

0.9 m/s (2.9 ft/s) with: 4.6 m (15 ft) vertical lift, 4.9 m (16 ft) of

3

/

8

-in. vinyl

intake tubing, 21 °C (70 °F) and 1524 m (5000 ft) elevation

Liquid sensor

Ultrasonic. Body: Ultem

®

NSF ANSI standard 51 approved, USP Class VI

compliant

Sample history

Up to 510 records

Air purge

Air purged automatically before and after each sample. The duration
automatically compensates for varying intake tube lengths.

Event log

510 records

Connections

Power, auxiliary, serial communications, distributor, SDI-12

Wetted materials

Stainless steel, polyethylene, Teflon, Ultem, silicone

Communications

RS232, Modbus, SDI-12

Warranty

1 year

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.