Table 3 pH and ORP sensor wiring – differential, 8350 and 8350.3/4/5 sensors with fixed cables
Terminal
Description
Sensor with an attached cable
Differential sensor
8350
8350.3/4/5
8-pin (J5)
1
Reference
Green
Black (coax shield)
Black (coax shield)
2
Ground solution
Clear
Jumper 1–2 on J5
Jumper 1–2 on J5
3
–V supply
White
—
—
4
—
—
—
—
5
—
—
—
—
6
Temp
Yellow
Red
Red
7
Temp/Circuit low
Black
White
White
8
—
—
—
2-pin (J4)
1
Active
Red
Transparent (coax core)
Transparent (coax core)
2
—
—
—
—
Sensor shield wires – Connect all sensor ground/shield wires to the controller enclosure
grounding screws.
Clear with a black band
—
Blue