Replacement parts and accessories – Hach-Lange POLYMETRON 9500 Contacting Conductivity Sensors User Manual User Manual

Page 8

Advertising
background image

1. Use a clean, soft cloth to remove loose debris from the end of the sensor. Rinse the sensor with

clean, warm water.

2. Soak the sensor for 2 to 3 minutes in a soap solution.
3. Use a soft bristle brush to scrub the entire measuring end of the sensor.
4. If debris remains, soak the measuring end of the sensor in a dilute acid solution such as < 5%

HCl for a maximum of 5 minutes.

5. Rinse the sensor with water and then return to the soap solution for 2 to 3 minutes.
6. Rinse the sensor with clean water.

Always calibrate the sensor after maintenance procedures.

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.

Sensors

Description

Item no.

2 electrode conductivity sensor K=0.01, ¾ inch NPT thread

08310=A=0000

2 electrode conductivity sensor K=0.1, ¾ inch NPT thread

08311=A=0000

2 electrode conductivity sensor K=1, ¾ inch NPT thread

08312=A=0000

2 electrode conductivity sensor K=0.01, ¾ inch NPT thread

08315=A=0000

2 electrode conductivity sensor K=0.01, for Yokogawa flow chamber

08315=A=0002

2 electrode conductivity sensor K=0.01, ¾ inch G thread

08315=A=1111

2 electrode conductivity sensor K=0.1, ¾ inch NPT thread

08316=A=0000

2 electrode conductivity sensor K=1, ¾ inch NPT thread

08317=A=0000

2 electrode conductivity sensor K=0.01, 1½ inch (38 mm) clamp

08394=A=1500

2 electrode conductivity sensor K=0.01, 1½ inch (38 mm) clamp with certificate of
conformity

08394=A=1511

2 electrode conductivity sensor K=0.01, 2 inch (51 mm) clamp

08394=A=2000

2 electrode conductivity sensor K=0.01, 2 inch (51 mm) clamp with certificate of
conformity

08394=A=2011

Cables

Description

Item no.

Female connector 6+T with connection drawing

08319=A=0000

5 m cable and IP65 connector for 2 electrode conductivity sensor

08319=A=0005

10 m cable and IP65 connector for 2 electrode conductivity sensor

08319=A=0010

20 m cable and IP65 connector for 2 electrode conductivity sensor

08319=A=0020

Shielded 4 conductor cable (per meter)

588800,29050

30 m cable and IP65 connector for 2 electrode conductivity sensor

91010=A=0144

8 English