Using the junction box, 8362 sensor – Hach-Lange POLYMETRON 9500 pH_ORP Sensors User Manual User Manual
Page 7

1 Model 2666 cable
9 -V - grey
2 Model 08350=A=8000 pre-amplifier
10 Input (measured pH or ORP) - green
3 Sensor
11 Temp+ (Pt100) - red
4 External shielding - EMC protection
12 Temp- (Pt100) - white
5 Temp+ (Pt100) - white
13 Ref. (reference) - black
6 Temp- (Pt100) - pink
14 Input (measured pH or ORP) - clear
7 Ref. (reference) - yellow
15 External shielding - EMC protection
8 +V - brown
Using the junction box
1 Model 2654 cable
9 External shielding - EMC protection
2 Model 08350=A=8500 junction box
10 External shielding - EMC protection
3 Sensor
11 Input (measured pH or ORP) - clear
4 Input (measured pH or ORP) - core
12 Ref. (reference) - black
5 Ref. (reference) - shielding
13 Temp+ (Pt100) - red
6 Temp+ (Pt100) - white
14 Temp- (Pt100) - white
7 Temp- (Pt100) - violet
15 Liquid earth (liquid contact) - blue
8 Solution ground (liquid contact) - yellow/green
For pH sensors only:
• Temperature measurement
• Impedance measurement (blue wire from the controller):
• Program the control frequency upper and lower limits (refer to the 9500 user manual):
• Glass impedance: 50 Mohm to 500 Mohm
• Reference impedance: 1 to 100 kohm
8362 sensor
The sensor is provided with two cables of 3 meters as standard, one for the pH sensor and one for
the temperature sensor. As an option, cables of 10 or 20 meters can be supplied. Attach the device
to a suitable static object using the U-bolt (No. 3 in
on page 5).
Follow these instructions to install the sensor (the numbers reference
English 7