Using the junction box, 8362 sensor – Hach-Lange POLYMETRON 9500 pH_ORP Sensors User Manual User Manual

Page 7

Advertising
background image

1 Model 2666 cable

9 -V - grey

2 Model 08350=A=8000 pre-amplifier

10 Input (measured pH or ORP) - green

3 Sensor

11 Temp+ (Pt100) - red

4 External shielding - EMC protection

12 Temp- (Pt100) - white

5 Temp+ (Pt100) - white

13 Ref. (reference) - black

6 Temp- (Pt100) - pink

14 Input (measured pH or ORP) - clear

7 Ref. (reference) - yellow

15 External shielding - EMC protection

8 +V - brown

Using the junction box

1 Model 2654 cable

9 External shielding - EMC protection

2 Model 08350=A=8500 junction box

10 External shielding - EMC protection

3 Sensor

11 Input (measured pH or ORP) - clear

4 Input (measured pH or ORP) - core

12 Ref. (reference) - black

5 Ref. (reference) - shielding

13 Temp+ (Pt100) - red

6 Temp+ (Pt100) - white

14 Temp- (Pt100) - white

7 Temp- (Pt100) - violet

15 Liquid earth (liquid contact) - blue

8 Solution ground (liquid contact) - yellow/green

For pH sensors only:

• Temperature measurement
• Impedance measurement (blue wire from the controller):

• Program the control frequency upper and lower limits (refer to the 9500 user manual):

• Glass impedance: 50 Mohm to 500 Mohm
• Reference impedance: 1 to 100 kohm

8362 sensor

The sensor is provided with two cables of 3 meters as standard, one for the pH sensor and one for
the temperature sensor. As an option, cables of 10 or 20 meters can be supplied. Attach the device
to a suitable static object using the U-bolt (No. 3 in

Figure 2

on page 5).

Follow these instructions to install the sensor (the numbers reference

Figure 2

on page 5):

English 7