4. Operation

4.5. Programming to measure ORP (redox). mV range

The instrument is supplied programmed to measure pH. To change the measuring unit (pH) to mV go to Setting-
up, after selecting the language, see page 12, or follow the diagram below until TYPE OF MEASURE.

Output current.
Selection of mV values correspon-
ding to 4 mA and 20 mA.
Selectable values between
± 1500 mV.

Language.
Selection of the working language:
English, German, Spanish, French,
Italian or Portuguese.

Hold.
Selection of the current output
when the instrument is in hold,
fixed at 21 mA or equivalent to
the last value

Variar con

CALIBRATE

CODE

000

LANGUAGE

Language > Español

Italiano

TYPE OF MEASURE

UNITS: pH
> mV

MEASURE pH & °C

M E A S U R E

pH 7.34 25.2 °C

OUTPUT 4-20 mA

4 mA 20 mA

-500 mV +500 mV

MEASURE

pH 7.34 25.2 °C

Introduce the code 100

STORING DATA

HOLD

> 21 mA

LAST

si627 P_HACH_LANGE_Ed.0510.indd ANG:14