7. Slowly raise the new cartridge to the measuring cell and screw into place to obtain an airtight and

watertight fitting.

8. Open the degassing valve (D).
9. Open the sample flow adjustment valve and ensure everything is watertight and there are no

leaks.

10. Close the degassing valve when the measuring cell is free of air.
11. Set the sample flow to the required rate (between 5 and 20 L/h).
12. Modify the resin parameters (see

Resin option

on page 11).

Resin replacement kit

1. Stop the sample flow using the sample flow adjustment valve (A).
2. Unscrew the used cartridge from the measuring cell and remove carefully by pulling down on the

cartridge until it clears the steel tube (B). The steel tube must remain in place.

3. Discard the used resin following the safety and disposal information for both the used resin and

used cartridges.

4. Fill the cartridge with the new resin up the marker line (C), adding demineralized water at regular

intervals to compact the resin.

5. Take the cartridge and place the end of the steel tube into the center (D).
6. Slowly raise the cartridge to the measuring cell and screw into place to obtain an airtight and

watertight fitting.

7. Open the degassing valve (E).
8. Open the sample flow adjustment valve and ensure everything is watertight and there are no

leaks.

9. Close the degassing valve when the measuring cell is free of air.

English 13