Approx. constant

0,6 cm

-1

Measuring range

10 µS/cm

to 80 mS/cm

Operating temp.

0 to 80 °C

Max. pressure

6 bar (25 °C)

Temperature sensor

Pt 1000

Body material

titanium

Electrode material

titanium

Thread

PG 13,5

Connector

MP-5

Min. immers. depth

10 mm

5

GB

Life of the cell
The cell lifespan is unlimited if suitable
maintenance is applied and the cell does
not break.

Troubleshooting
... Reading 0 independently of the mea-

sured sample.

Broken

cable.

Replace the cell.
... Unstable reading.
... Impossible calibration.
Dirty measuring electrodes or air

bubbles on their surface.

Clean or shake slightly the cell.
... The obtained readings differ a lot from

the real values.

Verify that the conductivity is covered

by the measuring interval of the cell.

Re-calibrate using standards in good

conditions.

... slow cell response
Regenerate the cell.

Specifications

Applications
Drinking waters, municipal waters, water
for irrigation, industrial waters, wastewa-
ter, etc.

CERTIFICATE OF QUALITY

The attached cell 53 88 has passed the
quality test and fulfils the following speci-
fications:
• EC error 5 %
between 10 µS/cm and 80 mS/cm.
• Temperature error 0.5 ºC.

53 88