4 replacing the sensor assembly, Maintenance – Hach-Lange 5740 sc User Manual
Page 24

20
Maintenance
5.4 Replacing the Sensor Assembly
The sensor assembly consists of a pre-installed membrane, electrolyte, and electrode.
1. Unscrew the sensor guard or optional lock ring and remove the spent or damaged
sensor assembly.
2. Install the o-ring onto the groove in the probe body, see item 2 and 3 in
Figure 5
.
3. Align the sensor assembly connector with the probe body socket, see item 6 in
Figure 5
.
4. Reinstall the sensor guard or lock ring.
5. Condition the new sensor before taking measurements. See
4.2 Sensor Setup on
page 11
.
Figure 5
Replacing the Sensor Assembly
1.
Probe Body
4.
Sensor Assembly
2.
Item 3 seats in groove.
5.
Sensor Guard (if sensor guard is not used, install the optional
lock ring)
3.
O-ring, (Cat. No. 1857100)
6.
Check for connector orientation during assembly.
2
6
3
1
4
5