4 replacing the sensor assembly, Maintenance – Hach-Lange 5740 sc User Manual

Page 24

Advertising
background image

20

Maintenance

5.4 Replacing the Sensor Assembly

The sensor assembly consists of a pre-installed membrane, electrolyte, and electrode.

1. Unscrew the sensor guard or optional lock ring and remove the spent or damaged

sensor assembly.

2. Install the o-ring onto the groove in the probe body, see item 2 and 3 in

Figure 5

.

3. Align the sensor assembly connector with the probe body socket, see item 6 in

Figure 5

.

4. Reinstall the sensor guard or lock ring.

5. Condition the new sensor before taking measurements. See

4.2 Sensor Setup on

page 11

.

Figure 5

Replacing the Sensor Assembly

1.

Probe Body

4.

Sensor Assembly

2.

Item 3 seats in groove.

5.

Sensor Guard (if sensor guard is not used, install the optional
lock ring)

3.

O-ring, (Cat. No. 1857100)

6.

Check for connector orientation during assembly.

2

6

3

1

4

5