Table 3 measurement options settings, Table 4 calibration options settings, Change calibration options – Hach-Lange HQD - Gel filled ORP_Redox electrode MTC User Manual

Page 11

Advertising
background image

DOC272.53.80033

11

.

6.

If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.

7.

Push

EXIT

until the meter returns to the measurement mode.

Change calibration options

1.

Make sure an electrode is connected to the meter.

2.

Push the

METER OPTIONS

key and select (MTC101) Settings.

3.

Select Modify Current Settings.

4.

Select Calibration Options.

5.

Update the settings. Refer to

Table 4

.

Table 3 Measurement options settings

Setting

Options

Description

Response time

Fast
Medium (default)
Slow

The response time affects the speed of the measurement by
adjusting the stabilization criteria.
Fast: 2 mV/min
Medium: 1 mV/min
Slow: 0.5 mV/min

Measurement limits

Lower limit (default: -1200.0
mV)
Upper limit (default: 1200.00
mV)

The measurement limits can be set to match the acceptable
values for the sample. When the measurement is above the
upper limit setting or below the lower limit setting, the meter
will show an "Out of limits" message.
This message is an alert to a potential problem with process
conditions.

Table 4 Calibration options settings

Settings

Option

Description

Standard

ZoBell’s (221mV -- 25ºC)
Light’s (468mV -- 25 ºC)
Custom

Temperature compensated for ZoBell’s
solution.
Standard values will appear on
Calibration Options screen.
Light’s solution is characterized at 25
ºC. Custom standard values and
temperature are user defined.

Calibration reminder

Reminder repeat: Off (default), 1 d, 7 d, 30 d

Meter will make an audible sound when
calibration is due. and repeat the sound
at selected interval.

Expires: Immediately, Reminder + 30 min,
Reminder + 1 hr, Reminder + 2 hr, Continue
Reading

Calibration expires after a specified
time.
Note: The meter cannot be used to read
samples after calibration has expired
unless Continue Reading is selected.