Hach-Lange HQD - Gel filled ORP_Redox electrode MTC User Manual

Page 4

Advertising
background image

4

DOC272.53.80033

Note: Procedures also apply for rugged model probes.

1.

Connect the

electrode to the
meter.

2.

Add fresh

ORP/Redox
standard to beaker
or appropriate
container.

3.

Select

Calibrate. The
display will show the
required
ORP/Redox
standard solution.

4.

Rinse the

electrode with
deionized water, blot
dry and place it in
the standard
solution. Make sure
that the sample
level is above the
reference junction.

5.

Push Read. Stir

gently. The display
will show
Stabilizing... and a
progress bar as the
probe stabilizes in
the standard.

6.

The display will

show the standard
solution value and
the mV offset when
the reading is
stable. Select Done
to view the
calibration
summary.

7.

The display will

show the calibration
summary. Select
Store to accept the
calibration and
return to the
measurement
mode.