Troubleshooting – Hach-Lange HQD - BOD LDO Probe LBOD User Manual

Page 10

Advertising
background image

10

DOC272.53.80025

Troubleshooting

Message or symptom

Possible cause

Action

Probe not supported

Software not updated

Update the HQd software to the newest revision at

www.hach.com/SoftwareDownloads.

Refer to the HQd

Series Meter manual.

HQd meter does not support IntelliCAL
probe.

Contact a Hach Technical Support Representative.

Connect a probe or
Probe requires service

Probe not connected properly

Disconnect, then connect the probe. Tighten the
locking nut.

Software not updated

Update the HQd software to the newest revision at

www.hach.com/SoftwareDownloads

Damaged probe

Verify connectivity with another probe or meter to
confirm isolated issue with probe. Contact a Hach
Technical Representative.

Out of range

Sensor cap scratched or damaged

Replace the LBOD sensor cap.

Large number of methods stored on
probe.

Continue to let probe connect. Do not disconnect
probe.

Temperature and/or pressure sensor
error.

Verify that both temperature and pressure sensors are
reading accurately.

1

Sample outside of specifications

Make sure that the sample measurement is within the
specifications of the probe or contact a Hach Technical
Representative.

Slope out of range

Probe not prepared for sample

Allow the probe to reach equilibrium in a selected
calibration environment (e.g. water saturated air) and
repeat calibration.

Calibration method settings

Verify the calibration standards in the method are
correct.

Sensor cap loose, scratched or damaged

Install or replace the sensor cap.

Temperature and pressure errors

Verify temperature and pressure are reading
accurately.

1

LED lights not functioning

Contact a Hach Technical Representative.

Stirrer assembly not
rotating

AC power not connected

Connect AC power

Switch in OFF position

Push the ON/OFF switch and verify that the indicator
light is illuminated.

Bent stirrer assembly

Remove the bearing clip and reposition the bearing
position in the bearing pocket. If ineffective, replace the
stirrer assembly.

Stirrer assembly noisy
(> 70 dB Tested 1 inch
from source in < 5 dB
background)

Bent stirrer assembly

Stirrer assembly contacting BOD
container.

Reposition the stirrer assembly in the BOD bottle or
replace the stirring assembly.

O

2

Sensor 0 days

remaining

There are 0 days remaining in the life of
the LBOD sensor cap.

Replace the LBOD sensor cap. Calibration will be
allowed, however the calibration icon along with a
question mark will appear on the measurement screen
even if the calibration has passed.

Meter currently set to incorrect date and
time.

Disconnect the probe from the meter. Set the correct
date and time in the Meter Options menu. Connect the
probe and verify that message has been removed.

Software not updated

Update the HQd software to the latest version and test
again.

O

2

Sensor ## of days

remaining

There are 30 days or fewer remaining in
the life of the LBOD sensor cap.

Replace the LBOD sensor cap soon.