Table 3 calibration options settings, Change calibration options – Hach-Lange HQD - BOD LDO Probe LBOD User Manual

Page 9

Advertising
background image

DOC272.53.80025

9

7.

If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.

8.

Push

EXIT

until the meter returns to the measurement mode.

Change calibration options

1.

Make sure a probe is connected to the meter.

2.

Push the

METER OPTIONS

key and select (Probe Model) Settings.

3.

Select Modify Current Settings.

4.

Select Calibration Options.

5.

Update the settings. Refer to

Table 3

.

6.

If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.

7.

Push

EXIT

until the meter returns to the measurement mode.

Table 3 Calibration options settings

Settings

Options

Description

Calibration

User — 100%
User — 100% with 0
User — mg/L
User — mg/L with 0 Resolution

Water-saturated air (100%) calibration

Calibration Reminder

Reminder Repeat On

8 h, 12 h, 1 d, 2 d, 5 d, 7 d

Expires

Immediately, Reminder + 30 min, Reminder +
1 h, Reminder + 2 h, Continue Reading