Table 3 calibration option settings, R to, Troubleshooting – Hach-Lange HQD - Luminescent Dissolved Oxygen Probe LDO User Manual

Page 11

Advertising
background image

DOC272.53.80021

11

6.

If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.

7.

Push

EXIT

until the meter returns to the measurement mode.

Table 3 Calibration option settings

Setting

Option

Description

Calibration

User—100%

Water-saturated air (100%) calibration

User—100% with 0

Water-saturated air (100%) calibration with 0 point

User—mg/L

Allows calibration with specified dissolved oxygen
concentration (mg/L) solution

User—mg/L with 0

Allows calibration with specified dissolved oxygen
concentration (m/L) solution with 0 point

Factory

Calibration with default LDO calibration

Standard value

Adjustable

Use arrow keys to specify the concentration of the solution
that will be used for calibration (if mg/L calibration mode is
selected).

Calibration reminder

Reminder Repeat

Off, 8 h, 12 h, 1 d, 2 d, 5 d, 7 d

Expires

Immediately, Reminder + 30 min, Reminder + 1 h, Reminder +
2 h, Continue Reading

Troubleshooting

Message or
symptom

Possible cause

Action

Probe not supported

Software not updated

Update the HQd software to the newest revision at

www.hach.com/SoftwareDownloads.

Refer to the HQd

Series meter manual.

HQd meter does not support IntelliCAL
probe

Contact a Hach Technical Support Representative.

Connect a probe or
Probe requires
service

Probe not connected properly

Disconnect, then connect the probe. Tighten the locking
nut.

Large number of methods stored on
probe

Continue to let probe connect. Do not disconnect the
probe. See the HQd Series meter manual.

Software not updated

Update the HQd software to the newest revision at

wwwhach.com/SoftwareDownloads

. Refer to the HQd

Series meter manual.

Damaged probe

Verify connectivity with another probe or meter to confirm
isolated issue with probe. Contact a Hach Technical
Representative.

Out of range

Sensor cap loose, scratched or
damaged

Reposition or replace the LDO sensor cap

Temperature and/or pressure sensor
error

Verify that both temperature and pressure sensors are
reading correctly

1

Damaged probe

Verify both blue and red LEDs are illuminated on the
probe. If not, replace the probe or contact a Hach
Technical Representative.

Sample outside of specifications

Verify that the sample concentration, temperature and
pressure are within the range of the LDO101 probe.

iButton number does not match sensor
cap lot number

Replace the iButton or sensor cap or perform a user
calibration

Sensor cap exposed to direct sunlight

Install the protective shroud