Hach-Lange HQD - Luminescent Dissolved Oxygen Probe LDO User Manual

Page 4

Advertising
background image

4

DOC272.53.80021

Note: Procedures also apply for rugged model probes.

1.

Connect the

probe to the meter.
The LDO101 probe
contains a factory
calibration. Remove
the shroud from the
probe.

2.

Push the

CALIBRATE

key to

create a new method
to enable user
calibration.

3.

Push the

METHODS

key, and

select User Cal -
100%. Push

OK

.

4.

Add

approximately ¼
inch (6.4 mm) of
reagent water to a
narrow-neck bottle,
such as a BOD
bottle.

5.

Insert a stopper

and shake
vigorously for
approximately 30
seconds to saturate
the entrapped air
with water. Allow up
to 30 minutes for
contents to
equilibrate to room
temperature.

6.

Remove the

stopper. Carefully
dry the cap with a
non-abrasive cloth
and place the probe
in the bottle.

7.

Select Read.

The display will
show Stabilizing...
and a progress bar
as the probe
stabilizes. The
display will highlight
the standard value.

8.

Select Done to

view the calibration
summary.

9.

Select Store to

accept the
calibration and
return to the
measurement mode.
Install the shroud on
the probe body.