Troubleshooting, If a calibration error occurs, refer to, If a measurement error occurs, refer to – Hach-Lange PHC28103 User Manual

Page 11

Advertising
background image

Note: The probe can be stored in a sample for up to 2 hours if the sample pH is not high.

1. Put the filling-hole cap in the filling hole (refer to

Figure 1

on page 2).

2. Rinse the probe with deionized water. Dry the probe with a lint-free cloth.
3. Fill the probe soaker bottle half full with 2.44 M KCl Electrode Filling Solution.
4. Loosen the soaker bottle cap and put the soaker bottle on the probe.
5. Turn the soaker bottle cap clockwise to tighten the soaker bottle cap.
6. Make sure that the solution in the soaker bottle completely covers the glass bulb and

reference junction holes.

If the glass bulb becomes dry:

1. Soak the probe tip in the 4.01, 7.00 and 10.01 buffers each for 5 minutes.
2. Rinse the probe with deionized water. Blot dry with a lint-free cloth.
3. Calibrate the probe.

Troubleshooting

Message or symptom

Possible cause

Action

Probe not supported

Software not updated

To download the most current version of the software,
refer to the applicable product page on the
manufacturer's website.
Refer to the HQd Series meter manual for specific
instructions for the meter model.

HQd meter does not support
IntelliCAL

®

probe

Contact a Technical Support Representative.

Connect a probe or probe
requires service

Probe not connected properly

Disconnect, then connect the probe. Tighten the locking
nut.

Software not updated

To download the most current version of the software,
refer to the applicable product page on the
manufacturer's website.
Refer to the HQd Series meter manual.

Large number of methods stored
on probe

Continue to let the probe connect. Do not disconnect the
probe.

Damaged probe

Make sure there is connectivity with another probe or
meter to confirm isolated issue with probe. Contact a
Technical Support Representative.

pH reading and/or mV
reading is same for all
solutions

Soaker bottle not removed

Remove the soaker bottle.

Electrical issue

Contact a Technical Support Representative.

Standard not recognized
error

Soaker bottle not removed

Remove the soaker bottle.

Incorrect or contaminated buffer
solution

Use fresh buffer solution as specified in the method.

Contaminated filling solution

Drain and replace the filling solution with fresh solution.

11