background image

4. Select Calibration Options and update the settings:

Option

Description

Buffer Set

Sets the temperature compensated buffer set used for calibration—

Color Coded—4.01, 7.00, 10.01 (default)

IUPAC—4.01, 7.00, 10.01, 12.45

DIN—4.65, 9.23

IUPAC—4.01, 6.86, 10.01, 12.45

IUPAC—4.01, 6.86, 9.18, 12.45

IUPAC—4.01, 7.00, 9.18, 12.45

Custom Buffer Set (refer to

Table 2

)

Custom buffer sets are characterized at 25 °C (77 °F).
Buffer set values are shown on the Calibration Options screen.
Note: Only the minimum calibration points must be measured for DONE to

be shown on the calibration screen.

Buffer Set
Values

If the Buffer Set is set to Customer Buffer Set, sets the custom buffer set
(refer to

Table 2

).

Minimum Cal
Points

Sets the minimum number of calibration points necessary before a
calibration can be completed—1 (default), 2 or 3.

Slope Limit

Sets the slope limit—1% to 10% (acceptable slope criteria, default = 5%).
The slope must fall within set limits for successful calibration.

5. Select Calibration Reminder and update the settings:

Option

Description

Reminder
Repeat

Meter will make an audible sound when a calibration is due and repeat the
sound at the selected interval—Off (default), 2 h, 4 h, 8 h, 2 d, 5 d or 7 d.

Expires

Calibration expires after the selected time—Immediately, Reminder +
30 min, Reminder + 1 h, Reminder + 2 h or Continue Reading.
Note: The meter cannot be used to read samples after calibration has

expired unless Continue Reading is selected.

6. If prompted, enter a name for the new method settings. Additional changes made to

the settings of an existing method are automatically saved with the same method
name.

7. Push EXIT until the meter returns to the measurement mode.