Specifications

Details

Storage solution

Saturated KCl

Minimum sample depth

14 mm (0.55 in.)

Dimensions

Body diameter: 12 mm (0.47 in.)
Body length: 103 mm (4.1 in.)
Total length: 200 mm (7.87 in.)
Cable length: 1 m (3.28 ft)

Cable connection

M12 digital output and connector compatible with HQd meters

1

Radiometer Analytical’s Red Rod technology: Encapsulation of reference elements with a red protection tube.

Product overview

The probe is a digital pH combination electrode with refillable electrolyte, a glass body
and an integrated temperature sensor. The probe is for laboratory use and has a high-
flow annular ring junction that measures pH in wastewater or "dirty" applications.

Note: Do not use the probe with organic solvents or high alkaline sample types.

A 59-mL bottle of reference electrolyte filling solution (saturated KCl solution) is included
with the probe. Refer to

Figure 1

.

Figure 1 Probe overview

1 Probe soaker bottle

5 Probe soaker bottle holder

2 Filling-hole cap

6 Reference junction

3 Protective tape and filling-hole

7 Glass bulb and temperature sensor

4 Cable, 1 meter (3.28 ft)

Preparation for use

Prepare the probe before a calibration or a sample measurement.

1. Turn the probe soaker bottle cap counter-clockwise to loosen the cap.
2. Remove the soaker bottle from the probe.
3. Fully rinse the reference junction and the glass bulb with deionized water to remove

all of the storage solution or possible salt particles. Dry with a no-lint cloth.

4. Remove the protective tape from the filling hole before initial use (refer to

Product

overview

on page 2). Discard the protective tape.