5. Select the Calibration Reminder and update the settings:

Option

Description

Reminder
Repeat

The meter makes a sound when a calibration is necessary and at the
selected interval—Off (default), 2 h, 4 h, 8 h, 2 d, 5 d or 7 d.

Expires

Calibration expires after the selected time—Immediately, Reminder +
30 min, Reminder + 1 h, Reminder + 2 h or Continue Reading.
Note: The meter cannot read samples after calibration has expired unless

Continue Reading is selected.

6. If prompted, enter a name for the new method settings. Other changes made to the

settings of an existing method are automatically saved with the same method name.

7. Push EXIT until the meter goes back to the measurement mode.

Table 3 Custom buffer sets

Buffer set
values

Option

Description

Std1 to
Std5

1.09 pH—25 °C Pre-set temperature compensated buffer values.

Note: It is necessary for selected standards to be different by a minimum of 2 pH units. For example, if
1.09 pH is selected for the first standard, the second standard must be different by at least 2 pH.
Standards that do not meet this minimum will show gray on the screen and will not be selectable.

1.68 pH—25 °C

4.01 pH—25 °C

4.65 pH—25 °C

6.86 pH—25 °C

7.00 pH—25 °C

9.18 pH—25 °C

9.23 pH—25 °C

10.01 pH—
25 °C

12.45 pH—
25 °C

Custom Buffer

Custom buffer value.
Range = 2.000 pH to 14.000 pH.
Custom buffer values are not temperature compensated. Custom buffers must be read at
25 °C (77 °C).

No Buffer

Select this option for undefined standards.

Change check standard options

1. Make sure that a probe is connected to the meter.
2. Push and select PHC735 Settings.
3. Select Modify Current Settings.