Specifications

Details

Storage solution

3 M KCl

Minimum sample depth

18 mm (0.71 in.)

Dimensions

Body diameter: 12 mm (0.47 in.)
Body length: 103 mm (4.1 in.)
Total length: 200 mm (7.87 in.)
Cable length: 1 m (3.28 ft)

Cable connection

M12 digital output and connector compatible with HQd meters

Product overview

The probe is a digital pH combination electrode with refillable electrolyte, a glass body
and an integrated temperature sensor. The probe is for laboratory use and measures pH
in general aqueous samples.

Note: Do not use the probe with organic solvents, low ionic strength, wastewater, high solids or

high alkaline sample types.

A 50-mL bottle of reference electrolyte filling solution (3 M KCl solution) is included with
the probe. Refer to

Figure 1

.

Figure 1 Probe overview

1 Probe soaker bottle

5 Probe soaker bottle holder

2 Filling-hole cap

6 Reference junction

3 Protective tape and filling-hole

7 Glass bulb and temperature sensor

4 Cable, 1 meter (3.28 ft)

Preparation for use

Prepare the probe before a calibration or a sample measurement.

1. Turn the probe soaker bottle cap counter-clockwise to loosen the cap.
2. Remove the soaker bottle from the probe.
3. Fully rinse the reference junction and the glass bulb with deionized water to remove

all of the storage solution or possible salt particles. Dry with a no-lint cloth.

4. Remove the protective tape from the filling hole before initial use (refer to

Product

overview

on page 2). Discard the protective tape.