calibrated again. To correct the probe calibration and status indicator, calibrate the
probe (refer to
Advanced operation
This section gives instructions about menu navigation and available options. Change
parameter-specific settings through the Full Access Options menu.
The settings that can be changed are shown in
Table 1 Parameter-specific settings
Setting
Options
Measurement Options
•
Resolution
•
Upper and lower range limits
Calibration Options
•
Buffer set
•
Buffer set values (if Custom Buffer Set selected)
•
Reminder
•
Minimum Cal points
•
Slope limit
Check Standard Options
•
Standard (temperature compensated buffer or custom at 25 °C (77 °F))
•
Reminder
•
Acceptance criteria
•
Standard value at 25 °C (77 °F) (if custom standard selected)
Units
•
pH
•
mV
Change measurement options
Methods are groups of default or user-defined settings related to specific applications. If
the meter is set to the default method, a prompt for a new name method shows. The
settings are saved with this name to identify them from the default method settings, which
cannot be changed. Use a saved method as an alternative of many adjustments to the
individual settings. Changes made to a user-defined method are automatically saved with
the existing name. Save different methods for the same probe on each meter.
1. Make sure that a probe is connected to the meter.
2. Push and select PHC805 Settings.
3. Select Modify Current Settings.
4. Select Units. Select pH (default) or mV.
Note: Use the mV option to find the probe offset in a pH 7 buffer or to measure the slope. Both
units show when the Detail Display mode is selected.
7