Troubleshooting, If a measurement error occurs, refer to – Hach-Lange PHC20103 User Manual User Manual

Page 10

Advertising
background image

2. Fill the probe storage cap or soaker bottle half full with Hach Electrode Storage Solution

or 3 M potassium chloride (KCl) solution.

3. Put the soaker bottle on the probe and tighten the soaker bottle cap.
4. Make sure that the solution in the storage cap or soaker bottle completely covers the

glass bulb and reference junction.

Note: The probe can also be stored in a sample for up to 2 hours if the sample pH is not high.

If the glass bulb becomes dry:
Rehydration

1. Soak the probe tip in the 4.01, 7.00 and 10.01 buffers each for 5 minutes.
2. Rinse the probe with deionized water. Blot dry with a lint-free cloth.
3. Calibrate the probe.

Troubleshooting

Message or symptom

Possible cause

Action

Probe not supported

Software not updated

Update the HQd software to the newest revision at

http://

www.hach.com/SoftwareDownloads

. Refer to the HQd Series

meter manual for specific instructions for the meter model.

HQd meter does not support
IntelliCAL

®

probe

Contact a Technical Support Representative.

Connect a probe or
probe requires service

Probe not connected properly

Disconnect, then connect the probe. Tighten the locking nut.

Software not updated

Update the HQd software to the newest revision at

http://

www.hach.com/SoftwareDownloads

. Refer to the HQd Series

meter manual.

Large number of methods stored

Continue to let probe connect. Do not disconnect the probe.

Damaged probe

Make sure there is connectivity with another probe or meter to
confirm isolated issue with probe. Contact a Technical Support
Representative.

Standard not recognized
error

Probe storage cap or soaker bottle
not removed.

Remove the probe storage cap or soaker bottle.

Incorrect or contaminated buffer
solution

Use fresh buffer solution as specified in the method.

pH reading and/or mV
reading is same for all
solutions

Storage cap or soaker bottle not
removed

Remove the storage cap or soaker bottle.

Electrical issue

Contact a Technical Support Representative.

Slow stabilization time

Contaminated glass sensor

Clean the probe (refer to

Clean the probe

on page 9).

Poor contact between reference
junction and solution

Shake the probe in the solution from side to side to refresh the
reference junction.

Probe not prepared for sample

For the best stabilization time, soak the probe in the sample
for several minutes before doing a sample measurement.

Low sample temperature or
temperature difference between
samples

Check the sample temperature. The lower the temperature or
the greater the difference of temperatures between samples,
the longer the stabilization time will be.

Air bubbles around inner reference
electrode

Gently tap the probe with hand or shake the probe downward
to remove any air bubbles.

10

DOC022.52.80197