Run a check standard

The run check standard feature validates instrument performance between sample
measurements. Use the run check standard feature for periodic or user-defined interval
measurements of a traceable standard solution. Set the criteria for check standards from
the PHC201 Settings menu.

Note: Access control must be off or a valid password must be entered before any of the check standard

method options can be changed.

1. Push . The Full Access Options menu is shown.
2. Select Run Check Standard.

Note: Select the correct probe if two probes are connected to the meter.

3. Get the standard solution shown on the display.
4. Rinse the probe with deionized water. Blot dry with a lint-free cloth.
5. Put the probe in the standard solution until the temperature sensor is completely

submerged. Move the probe up or down or gently shake the probe to remove air
bubbles.

6. Push Read. The display will show "Stabilizing" and a progress bar as the reading

stabilizes. The display shows the value of the check standard and either Check
Standard Passed or Check Standard Failed.

7. If the display shows Check Standard Passed, the check standard measurement is

within the accepted limits set by the administrative user. Select Done to continue with
the sample measurement.

8. If the display shows Check Standard Failed, the measurement is outside of accepted

limits set by the administrative user and a recalibration is recommended. If the
acceptance criteria is set to Cal Expires on Failure: Yes, the display shows the
calibration icon and a question mark until the probe is recalibrated. To correct the probe
calibration and status indicator, calibrate the probe (refer to

Calibration

on page 2).

Advanced operation

Parameter-specific settings can be changed through the Full Access Options menu. Details
about menu navigation, available options and how to change them are given in the screens,
tables and procedures throughout this section.
The settings that can be changed are shown in

Table 1

.

Table 1 Parameter-specific settings

Setting

Options

Measurement Options

Resolution

Upper and lower range limits

Calibration Options

Buffer set

Buffer set values (if Custom Buffer Set selected)

Reminder

Minimum Cal points

Slope limit

Check Standard Options

Standard (temperature compensated buffer or custom at 25 °C)

Reminder

Acceptance criteria

Standard value at 25 °C (if custom standard selected)

Units

pH

mV

DOC022.52.80197

5