6. Select Acceptance Criteria and update the settings:

Option

Description

Acceptance Limits

Sets the tolerance limits for check standard—1% to 20%.

Cal Expires on Failure Recalibration required if check standard fails—Yes or No.

The calibration expires if the check standard fails and Cal Expires is
set to Yes.

7. If prompted, enter a name for the new method settings. Additional changes made to

the settings of an existing method are automatically saved with the same method
name.

8. Push EXIT until the meter returns to the measurement mode.

Maintenance

Clean the probe

Clean the probe when:

Drifting/inaccurate readings occur as a result of contamination on the sensing
element or improper storage conditions.

Slow response time occurs as a result of contamination on the sensing element.

The slope is out of range as a result of contamination on the sensing element.

For general contaminants, complete the following steps.

1. Rinse the probe with deionized water. Blot dry with a lint-free cloth. Do not touch the

tip of the probe.

2. If harsh contaminants are attached to the probe, polish the probe tip with a soft cloth

or cotton swab to remove the contaminants.

3. Soak for 30 minutes in 100 mg/L Nitrate Nitrogen standard solution.

Storage

The probe can be stored dry. To protect the sensing element, rinse with DI water and blot
dry with a lint-free cloth. Install the sensor protection cap.

Note: The probe must be conditioned after long-term storage. Refer to

Preparation for use

on page 2.

Troubleshooting

Message or symptom

Possible cause

Action

Probe not supported

Software not updated

To download the most current version of the
software, refer to the applicable product page on the
manufacturer's website.
Refer to the HQd Series meter manual for the
specific instructions for the meter model.

HQd meter does not support
IntelliCAL

®

probe

Contact a Technical Support Representative.