Message or symptom

Possible cause

Action

Connect a probe or probe
requires service

Probe not connected properly

Disconnect, then connect the probe. Tighten the
locking nut.

Software not updated

To download the most current version of the
software, refer to the applicable product page on the
manufacturer's website.
Refer to the HQd Series meter manual.

Large number of methods stored on
probe.

Continue to let probe connect. Do not disconnect
probe.

Damaged probe

Make sure connectivity with another electrode or
meter to confirm isolated issue with probe. Contact a
Technical Support Representative.

mV reading is the same for all
solutions

Electrical issue

Contact a Technical Support Representative.

Slow response time

Dirty sensing element

Clean the probe (refer to

Clean the probe

on page 10).

Low sample temperature or
temperature difference between
samples

Check the sample temperature. The lower the
temperature or the greater the difference of
temperatures between samples, traditionally the
longer the response time.

Bubbles trapped under sensor tip

Gently shake the probe until bubbles are removed
from under sensor tip.

Slope out of range (refer to

Check probe response

on page 13)

pH is incorrect

Make sure the pH is between 3 and 5 after each ISA
addition.

Ionic strength adjustor (ISA) not
used

Add ISA to each sample and standard (one powder
pillow per 25 mL of solution).

Insufficient conditioning

Condition for at least 30 minutes in a 100 mg/L
standard solution with ISA.

Damaged probe

Contact a Technical Support Representative.

Incorrect standards

Calibrate using freshly prepared standards.

Dirty sensing element

Clean the probe and recalibrate.

Bubbles trapped under sensor tip

Gently shake the probe until bubbles are removed
from under sensor tip.