5.5. Readjusting temperature
The instrument can correct the deviation of a temperature
probe (built into pH electrodes) at 25 °C and 85 °C (77 °F
and 185 °F).
This allows the instrument to be used as a precision thermometer.
Follow the diagram.
2 s
Enter the correction of
the temperature sensor at
25 °C (77 °F).
Enter the correction of
the temperature sensor at
85 °C (185 °F).