Hach-Lange PHC10130 User Manual User Manual

Page 10

Advertising
background image

Message or symptom

Possible cause

Action

Drifting/Inaccurate readings

Incorrect settings

Measurement Options—Make sure that Temp Correction
(Correction Factor if not set as NaCl) and reference
temperature are both correct.

Calibration Options—Make sure that Standard Value,
Reference Temperature and Temp Correction are all
correct.

Mineral or sample buildup on
cell

Clean the probe (refer to

Clean the probe

on page 8).

CO2 absorption in LIS/high
purity samples

Isolate LIS/high purity samples to prevent sample
contamination.

Bubbles trapped under probe
tip

Gently shake the probe until bubbles are removed.

Calibration failed - Out of
limits/Out of range

Incorrect settings

Measurement Options—Make sure that Temp Correction
(Correction Factor if not set as NaCl) and reference
temperature are both correct.

Calibration Options—Make sure that Standard Value,
Reference Temperature and Temp Correction are all
correct.