Install the shroud

1. Put the locking ring on the probe with the threads toward the probe.
2. Slide the shroud on the probe until it is against the locking groove.
3. Hand-tighten the locking ring on the shroud.

Storage

Between uses, make sure the probe is dry and store it in ambient conditions. Rugged
probes may be stored with the shroud installed if the storage container is sufficiently
large.

Troubleshooting

Message or symptom

Possible cause

Action

Probe not supported

Software not updated

To download the most current version of the software, refer
to the applicable product page on the manufacturer's
website.
Refer to the HQd Series meter manual for specific
instructions for the meter model.

HQd meter does not support
IntelliCAL

®

probe

Contact a Technical Support Representative.

Connect a probe or probe
requires service

Probe not connected
correctly

Disconnect, then connect the probe. Tighten the locking
nut.

Software not updated

To download the most current version of the software, refer
to the applicable product page on the manufacturer's
website.
Refer to the HQd Series meter manual.

Large number of methods
stored on the probe

Continue to let probe connect. Do not disconnect the probe.

Damaged probe

Make sure there is connectivity with another probe or meter
to confirm isolated issue with probe. Contact a Technical
Support Representative.

Slow stabilization time

Mineral or sample buildup on
electrodes

Clean the probe (refer to

Clean the probe

on page 8).

Bubbles trapped under probe
tip

Make sure that the sample concentration and temperature
are both within the range of the CDC401 probe.

Out of range

Temperature and/or pressure
sensor error

Make sure that the temperature and pressure sensors are
both reading correctly.

Damaged probe

Replace the conductivity probe or contact a Technical
Support Representative.

CO2 absorption in LIS/high
purity samples

Replace the conductivity probe or contact a Technical
Support Representative.

Bubbles trapped under probe
tip

Make sure that the sample concentration and temperature
are both within the range of the CDC401 probe.

9