Table 2 Error code descriptions (continued)

Error code

Description

Solution

E13

The ISFET pH probe temperature
sensor is out of range.

Replace the probe. Contact
technical support for probe
replacement information.

E20

The conductivity probe temperature
sensor is out of range.

Replace the probe. Contact
technical support for probe
replacement information.

Replacement parts and accessories

W A R N I N G

Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.

Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.

Replacement parts

Description

Quantity

Item no.

AA Alkaline batteries

4

1938004

Accessories

Description

Quantity

Item no.

ISFET pH probes

General purpose round tip probe

1

PH77-SS

Micro pH probe

1

PH17-SS

Piercing tip micro pH probe

1

PH37-SS

pH probe for NMR tubes

1

PH47-SS

Heavy duty piercing tip micro pH probe

1

PH57-SS

Conductivity probes

General purpose conductivity probe, epoxy, K = 0.5

4 graphite contacts

CDW97-KP5

LIS conductivity probe, glass, K = 1

4 platinum contacts

CDW97-K1

Conductivity probe for salinity and brackish water, epoxy, K =
10

4 contacts

CDW97-K10

Glass pH probes

Gel-filled combination pH, BNC and 3.5 mm phone

1

5193511

Pt Series combination pH, BNC and 3.5 mm phone

1

5191011

Refillable combination pH, BNC and 3.5 mm phone

1

5194011

English 19

Advertising