3

LDO101-05, LDO101-10, LDO101-15, or LDO101-30 Probe

Replacing the Shroud

1. Place the locking ring (

Figure 2

, item

2

) on the probe body with the threads toward the

shroud (

Figure 2

, item

1

).

2. Slide the shroud onto the probe until the locking ribs (

Figure 2

, item

4

) are seated in

the locking groove (

Figure 2

, item

7

).

3. Screw the locking ring onto the threads.

4. Hold the shroud and hand-tighten the locking ring.

Calibration

Each lot of LDO sensor caps is factory-calibrated. For best performance, a one-time
calibration initialization can be performed when a new sensor is installed. Additional
calibrations can be performed but are not required.

For calibration steps, refer to the HQ Portable Meters User Manual. The IntelliCAL™
probes store the current calibration and calibration history in the probe memory. It is not
necessary to recalibrate when moving a calibrated probe from one meter to another.

Figure 2 Shroud and Sensor Cap Removal

1

Shroud

6

Cap Seal

2

Locking Ring

7

Locking Groove

3

Sensor Cap

8

Temperature Sensor

4

Locking Ribs (4x)

9

Sensor Lens

5

O-ring