Hach-Lange PHC30103 User Manual User Manual

Page 11

Advertising
background image

Message or symptom

Possible cause

Action

Connect a probe or probe
requires service

Probe not connected properly

Disconnect, then connect the probe. Tighten the locking
nut.

Software not updated

To download the most current version of the software,
refer to the applicable product page on the
manufacturer's website.
Refer to the HQd Series meter manual.

Large number of methods stored
on probe

Continue to let the probe connect. Do not disconnect the
probe.

Damaged probe

Make sure there is connectivity with another probe or
meter to confirm isolated issue with probe. Contact a
Technical Support Representative.

pH reading and/or mV
reading is same for all
solutions

Soaker bottle not removed

Remove the soaker bottle.

Electrical issue

Contact a Technical Support Representative.

Standard not recognized
error

Soaker bottle not removed

Remove the soaker bottle.

Incorrect or contaminated buffer
solution

Use fresh buffer solution as specified in the method.

Contaminated filling solution

Drain and replace the filling solution with fresh solution.

Slow stabilization time

Tape not removed from the filling-
hole

Remove the tape that is over the filling hole.

Contaminated glass sensor

Clean the probe (refer to

Clean the probe

on page 9).

Contaminated filling solution

Drain and replace the filling solution with fresh solution.

Filling hole closed

Open the filling hole while in use by removing the filling-
hole cap from the filling hole.

Poor contact between reference
junction and solution

Shake the probe in the solution from side to side to
refresh the reference junction.

Probe not prepared for LIS sample For the best stabilization time, soak the probe in the

sample for 10 to 15 minutes before the initial
measurement.

Low sample temperature or
temperature difference between
samples

Check the sample temperature. The lower the
temperature or the greater the difference of temperatures
between samples, the longer the stabilization time will
be.

Air bubbles around inner
reference electrode

Gently tap the probe with hand or shake the probe
downward to remove any air bubbles.

Calibration errors

Calibration not done correctly

Recalibrate using freshly prepared pH buffers.

Contaminated glass sensor

Clean the probe (refer to

Clean the probe

on page 9).

Slope exceeds the criteria for % of
theoretical (as defined in the
method slope limit)

1.

Widen the slope limits by changing the PHC301
calibration settings and method.

2.

Recalibrate the probe.

3.

Run a check standard to check the performance of
the probe.

11