Preparation for use – Hach-Lange PHC30103 User Manual User Manual

Page 2

Advertising
background image

Figure 1 Probe overview

1 Probe soaker bottle

4 Protective tape and filling-hole

2 Glass bulb, reference junction and temperature sensor

5 1 or 3 meter (3.28 or 9.84 ft) cable

3 Filling-hole cap

6 Probe soaker bottle holder

Preparation for use

Prepare the probe for use before calibration or sample measurement.

1. Turn the probe soaker bottle cap counter-clockwise to loosen the cap.
2. Remove the soaker bottle from the probe.
3. Rinse the reference junctions and glass bulb with deionized water thoroughly to

remove the viscous storage/filling solution completely. Blot dry with a lint-free cloth.

4. Remove the protective tape from the filling hole before initial use (refer to

Figure 1

on page 2). Dispose of the protective tape.

5. Add filling solution to the probe as necessary (refer to

Fill the probe

on page 10).

The filling solution must be above the standard solution or sample level during
measurement (

Figure 2

).

6. Make sure that the filling hole is open during measurement for the proper flow of the

filling solution.

Note: For the best stabilization time, condition the probe for several minutes in the sample

before use.

Figure 2 Measurement method

1 Container

3 Filling solution level

2 Probe body

4 Standard solution or sample level

2

Advertising
This manual is related to the following products: