View the calibration data, Set the calibration reminder, Sample measurements – Hach-Lange SENSION+ MM374 User Manual

Page 11

Advertising
background image

8. Push to measure the second calibration solution.

The next calibration solution is shown.

9. Rinse the probe with deionized water and put the probe into the third

calibration beaker. Be sure that there are no air bubbles in the
membrane.

10. Push to measure the third calibration solution.

When the calibration is good, the display briefly shows Calibration
OK and then returns to the main menu.

Note: When a printer is connected the print menu opens and the result can be
printed.

View the calibration data

Data from the most recent calibration can be shown.

1. From the main menu use the or to select DATA LOGGER.

Confirm.

2. Select Display data.
3. Select Calibration data and confirm with . The last calibration data

is shown.

• pH-the slope and offset values are shown alternating with the

deviation (in %) and calibration temperature.

• ORP-the measured mV value and calibration temperature are

shown.

• Conductivity-the cell constant and calibration temperature for each

standard are shown.

Set the calibration reminder

pH: The calibration reminder can be set between 0 to 23 hours or
1-7 days (default 1 day). EC: The calibration reminder can be set
between 0 to 99 days (default 15 days). The display shows the
remaining time to the new calibration.

Note: When 0 days is selected, the calibration reminder is turned off.

1. From the main menu use the or to select CALIBRATION.

Confirm.

2. Use the to enter the calibration menu.
3. Use the or to select Cal. frequency and confirm.

4. Use the and to advance to the next step and use the or to

change a value. Confirm.
Push to start the calibration.

Sample measurements

Each probe has specific preparation steps and procedures for taking
sample measurements.

1. From the main menu use the or and and to select

MEASURE. Confirm.

2. Use the to change the following settings. Confirm every entry.

Option

Description

Resolution

Select the resolution: 1, 0.1, 0.01 (default) or 0.001

Measure

Stability—select By stability Criterion: Fast (variation <
0.02 pH in 6 s), Standard (variation < 0.01 pH in 6 s) or Strict
(variation < 0.002 pH in 6 s). In continuous—enter the time
interval for the In continuous Acquis. interval (data storage or
printing data). By time—enter the time interval for data storage
or printing data.

Display mV Display mV—select YES or NO to show mV.

Limits

Limits—select YES or NO. YES: Enter the upper and lower
limits. An acoustic warning appears when the measurement is
out of limit. The report output shows an A next to the
measured value when the measurement was out of limit.

Isopotential Isopotential—change the Isopotential pH value in Data

introduction. Select Calculate to calibrate the probe again.

TC

TC—select Linear or Natural waters. Linear: enter a value in
%/°C (default 2.00%/Temperature). Natural waters: Non-linear
for natural waters, according to EN27888

Tref

Reference temperature—select between 20 or 25 °C or Other
temperature.

3. Push to start the measurement.

Note: If the measurement is not stabilizing after 120 seconds, the meter turns
automatically into the continuous measurement mode.

English 11