Using the pour-thru cell, Dr5000 pour-thru cell kit installation – Hach-Lange DR 5000 Pour-Thru Cell Kit Installation User Manual

Page 7

Advertising
background image

7

DR5000 Pour-Thru Cell Kit Installation

Using the Pour-Thru Cell

Most HACH methods prescribe the use of round cells.

When a pour-thru cell is used, the measurement results of these HACH methods must be
multiplied by a dilution factor of 0.95 or revised with the help of a standard adjustment. The
appropriately revised method can be stored as a user program. If the use of a pour-thru
cell is prescribed in a HACH method, no adjustment is needed.

A pour-thru cell can be used for a HACH method if the prescribed sample volume is 25 ml.

Although the pour-thru cell can be used for the following HACH methods, the cell must be
thoroughly purged with distilled water between individual samples.

The Pour-Thru Cell can also be used to run Nitrogen, Ammonia, Nessler Method and TKN
chemistries. Clean the cell by pouring a few sodium thiosulfate pentahydrate crystals into
the cell. Rinse out the crystals with deionized water.

WARNING:
The pour-through cell cannot be used for the following HACH methods.

Table 1 HACH methods for which additional purging of the pour-through cell is

necessary

Aluminum, Aluminon

Chlorine Dioxide, LR

Cobalt, PAN

Copper, Porphyrin

Hardness, Calmagite

Manganese, LR, PAN

Nickel, PAN

Nitrate, MR

Nitrate, HR

Table 2 HACH methods for which the pour-through cell cannot be used

Aluminum ECR

Arsenic

Barium

Boron, Carmine

Cyanuric Acid

Fluoride

Formaldehyde

Lead, LeadTrak

Mercury

Nickel, Heptoxime

Nitrite, HR

PCB

Phenols

Potassium

Selenium

Silver

Suspended Solids

Sulfate

TPH

Volatile Acids

Zinc

Surfactants, Anionic (Detergents)