4 sensor diagnostics menu, 5 sensor setup menu, 4 sensor diagnostics menu 5.5 sensor setup menu – Hach-Lange NITRATAX sc User Manual

Page 20

Advertising
background image

18

5.4 Sensor diagnostics menu

SELECT SENSOR STATUS>SELECT SENSOR (if more than one sensor is attached)

ERROR LIST

Displays all actual error messages:
MOIST, R < M, DEXT < 0.0, W. POS. UNKNOWN, W. BLOCKED, FLASH FAILURE, R TOO
HIGH, REPLACE SHAFT SEAL, SENSOR MISSING

WARNING LIST

Displays all actual error messages:
EM TOO HIGH, CONC. TOO HIGH, CHECK CALIBR., REPLACE PROFILE, SERVICE
REQUIRED, REPLACE SEALS, SHAFTSEALS REPL.

Note: For more information about error messages and warnings refer to

Section 7 on page 29

.

5.5 Sensor setup menu

SELECT SENSOR (if more than one sensor is attached)

CALIBRATE (see

5.6 on page 20

)

FACTOR

Correction factor for the measured value.
Possible settings: 0.80–1.20
Default: FACTOR = 1

OFFSET

Adjustable from –250 to +250 mE for zero point correction
Default: OFFSET = 0

OFFSET ADJUST

Perform a zero point calibration

1 SAMPLE CAL

Perform a single point calibration

CAL CONFIG

Select OUTPUT MODE or CAL INTERVAL
OUTPUT MODE: Select the behavior of the outputs during calibration for zero point setting
(Hold, Active, Transfer, Choice). Hold maintains the last reading prior to going into the menu.
Active transmits the current level readings, corrected with previous calibration data until new
data is entered. Set Transfer transmits the value designated during the system setup
CAL INTERVAL: Enter number of days

SET CAL DEFLT

The instrument resets the settings to the default configuration.

CONFIGURATION

EDIT NAME

Can be edited as required (up to 10 characters)

PARAMETER

NOx-N or NO3 (eco only NOx-N)

MEAS UNIT

Unit for the measured result.
Possible settings: mg/l, ppm

MEAS INTERVAL

eco/clear: 5, 6, 10, 12, 15, 20, 30 min
plus: 15, 20, 30 sec; 1, 2, 3, 4, 5, 6, 10, 12, 15, 20, 30 min
Note: Intervall of the data log is the same like the measuring intervall.

RESPONSE TIME

Indication of the actual response time in Counts (count x measuring interval = response time)
eco: 3–6 x MEAS INTERVAL
clear: 1–6 x MEAS INTERVAL
plus: 1–12 x MEAS INTERVAL
Note: Gliding average over 2-12 measurements.

CLEANING

eco, clear: 1/MEASURE
plus: 1/MEASURE; 1,2,3,5,6,10,12,15,20,30 min; 1,2,3,4,6,12 h, 10:00 h

WIPER MODE

Wiping interval.
Select SINGLE or DOUBLE A-B-A or DOUBLE B-A-B
SINGLE: Default setting (Default: eco)
DOUBLE A-B-A: Double wiping frequency
DOUBLE B-A-B: Double wiping frequency (Default: plus, clear)

BYPASS

YES/NO (plus and clear)
YES: Setting for bypass application (inhibits wiper "extension")

Advertising