Hach-Lange NITRATAX sc User Manual

Page 5

Advertising
background image

3

Section 1

Specifications

Specifications are subject to change without notice.

Component

NITRATAX plus sc

NITRATAX eco sc

NITRATAX clear sc

NITRATAX sc tank sensor

Measuring technique

UV absorption measurement, reagent-free

Measuring method

Patented 2-beam method

Measuring path

1 mm (0.04 in.), 2 mm
(0.08 in.), 5 mm (0.20 in.)

1 mm (0.04 in.)

5 mm (0.20 in.)

Measuring range with
NO

3

-N standard solutions

0.1-100.0 mg/L NO

2+3

–N

(1 mm/0.04 in.)
0.1–50.0 mg/L NO

2+3

–N

(2 mm/0.08 in.)
0.1–25.0 mg/L NO

2+3

–N

(5 mm/0.20 in.)

1.0–20.0 mg/L NO

2+3

–N

0.5–20.0 mg/L NO

2+3

–N

Lower detection limit
(mg/L) NO

3

-N

0.1 (5 mm/0.20 in.)

1

0.5

Upper detection limit
(mg/L) NO

3

-N

100 (1 mm/0.04 in.)

20

20

Measuring error (mg/L)
NO

3

-N

±3 % of the mean MW ±0.5

±5 % of the mean MW ±1.0

±5 % of the mean
MW ±0.5

Resolution (mg/L)

0.1

0.5

0.1

Sludge compensation

yes

yes

Measuring interval (>= min) 1

5

5

T100 response time (min)

1

15

5

Integration

>1 min, adjustable

15–30 min, adjustable

>5 min, adjustable

Power consumption

2 W

Cable length

10 m (30 ft)

Sensor pressure limit

maximum 0.5 bar (7 psi)

Ambient temperature

2 to 40 °C (36 to 100 °F)

Dimensions D x L
(

Figure 1 on page 5

)

approximately
70 x 229–333 mm
(3 x 13.1 in.)

approximately
75 x 323 mm
(3 x 12.9 in.)

approximately
75 x 327 mm
(3 x 12.7 in.)

Weight

approximately 3.6 kg (7.9 lb)

approximately 3.3 kg (7.3 lb)

approximately 3.3 kg (7.3 lb)

NITRATAX sc sensors flow through units

Sample flow rate

0.5–10 L/h sample

0.5–10 L/h sample

Sample connection

Tube ID 4 mm/AD 6 mm

Tube ID 4 mm/AD 6 mm

Sample temperature

2 to 40 °C (36 to 100 °F)

2 to 40 °C (36 to 100 °F)

Dimensions

W x H x D approximately
500 x 210 x 160 mm
(20 x 8.3 x 6.3 in.)

W x H x D approximately
500 x 210 x 160 mm
(20 x 8.3 x 6.3 in.)

Weight (without sensor)

approximately 3.6 kg (7.9 lb)

approximately 3.6 kg (7.9 lb)

Advertising