LDO - Changing sensor cap
•
Change the sensor cap once a year or if the measurements are excessively
inaccurate despite cleaning and calibration.
•
Do not expose the inside of the sensor cap to direct or indirect sunlight.
•
Leave the sensor cap in the packaging until immediately prior to installation.
•
Enter the sensor code as described below.
Fig. 1
Changing sealing ring and sensor cap
1.
Clean the seal groove (1) before you fit a new sealing ring (2).
2.
Place the sealing ring in the seal groove.
3.
Fit and hand-tighten the new sensor cap (3).
1
2
3