Chain Mount Kit LZX914.99.12400

Selecting the installation location

Important note: Refer to the sensor user manual when installing the mounting hardware.
Note all safety instructions and hazards.

Note: When selecting the installation location, always ensure there is safe access to all parts of the
system and that the sensor can never hit the tank wall.

During the selection of the installation location, pay attention to:

The load bearing capacity and strength of the ground or railings (leverage due to flow
and installation)

The maximum cable and tubing lengths

The flow conditions

The space required to remove the sensor for cleaning

Note: Refer to the sensor user manual for sensor-specific installation requirements.

Figure 2 Installation dimensions