Clean the power connection socket, Replace the fuse – Hach-Lange Power connection box for SC analyzers User Manual

Page 10

Advertising
background image

10

Figure 8

Circuit diagram (see inside of the lid)

Clean the power connection socket

Wipe the outside of the housing with a damp cloth. Make sure that the connector is well
sealed with the sealing cap. Use a mild cleaning agent if required. Solvents must not be
used.

Replace the fuse

1. Make sure that the power connection box is disconnected from the power supply.

1

Phase, line (live)

7

Black for left power socket

2

Connection to terminal 1 via bridge

8

Black for right power socket

3

Connection to terminal 4 via bridge

9

White for left power socket

4

Neutral conductor

10 White for right power socket

5

Protective ground (two terminals)

11 Fuse, T 5 A H, 250 V (4x)

6

Protective ground for left and right power socket

12 Lid protective ground

D A N G E R

Risk of electric shock.
Disconnect the power connection socket from the power supply before it is cleaned in order to
avoid electric shocks.

D A N G E R

Risk of electric shock.
The procedures described in this section of the instruction sheet may only be performed by
qualified personnel.

D A N G E R

Risk of electric shock.
An incorrect fuse can lead to injuries and damage. The fuse must always be replaced with a fuse
of the same type and with the same nominal values.