2 data table – Hach-Lange PCII-COD LCK 014_114_314_614 User Manual
Page 24

System start up
22
•
high range with POCKET COLORIMETER II 600 nm
•
LCK 014: 1000-10000 mg/l COD
•
LCK 114: 150-1000 mg/l COD
•
low range with POCKET COLORIMETER II 450 nm
•
LCK 314: 15-150 mg/l COD
•
LCK 614: 50-300 mg/l COD
Each programming procedure has only to be followed once prior to the first test
evaluation.
For detailed information please read the POCKET COLORIMETER II instruction
manual.
3.2.2 Data table
3.2.2.1
Evaluation COD (high range) with POCKET COLORIMETER II 600 nm
3.2.2.2
Evaluation COD (low range) with POCKET COLORIMETER II 450 nm
Table 1 Evaluation COD (high range) with POCKET COLORIMETER II 600 nm
LCK 014
LCK 114
RES (resolution)
----
----
S0 (standard 0)
1000
0150
A0 (absorbance 0)
0.177
0.075
S1 (standard 1)
9999
1000
A1 (absorbance 1)
1.650
0.505
Table 2 Evaluation COD (low range) with POCKET COLORIMETER II 450 nm
LCK 314
LCK 614
RES (resolution)
---.-
---.-
S0 (standard 0)
015.0
050.0
A0 (absorbance 0)
-0.062
-0.195
S1 (standard 1)
150.0
300.0
A1 (absorbance 1)
-0.593
-1.15