System start up

22

high range with POCKET COLORIMETER II 600 nm

LCK 014: 1000-10000 mg/l COD

LCK 114: 150-1000 mg/l COD

low range with POCKET COLORIMETER II 450 nm

LCK 314: 15-150 mg/l COD

LCK 614: 50-300 mg/l COD

Each programming procedure has only to be followed once prior to the first test

evaluation.

For detailed information please read the POCKET COLORIMETER II instruction

manual.

3.2.2 Data table

3.2.2.1

Evaluation COD (high range) with POCKET COLORIMETER II 600 nm

3.2.2.2

Evaluation COD (low range) with POCKET COLORIMETER II 450 nm

Table 1 Evaluation COD (high range) with POCKET COLORIMETER II 600 nm

LCK 014

LCK 114

RES (resolution)

----

----

S0 (standard 0)

1000

0150

A0 (absorbance 0)

0.177

0.075

S1 (standard 1)

9999

1000

A1 (absorbance 1)

1.650

0.505

Table 2 Evaluation COD (low range) with POCKET COLORIMETER II 450 nm

LCK 314

LCK 614

RES (resolution)

---.-

---.-

S0 (standard 0)

015.0

050.0

A0 (absorbance 0)

-0.062

-0.195

S1 (standard 1)

150.0

300.0

A1 (absorbance 1)

-0.593

-1.15