Section 4

Operation

4.1 Analyse COD with LANGE Cuvette Tests

DANGER

Handling chemical samples, standards and reagents can be dangerous. Users

of this product are advised to familiarize themselves with safety procedures

and the correct use of chemicals, and to carefully read all relevant material

safety data sheets.

To analyse COD follow the working procedure enclosed in the according Cuvette Test

LCK 014/114/314/614.

4.2 Quality assurance

The measurement results must be subjected to plausibility checks (dilute and/or

spike the sample).

Regular checks with corresponding standard solutions (see

Section 5 Replacement

Parts and Accessories on page 31

) ensure that your measurement system is

functioning properly and is being correctly operated.

4.3 Evaluate LANGE COD Cuvette Tests

Note: Make sure that the instrument is programmed with the corresponding calibration data,
see

3.4 Check entered data on page 28

.

1.

Preparation of a reagent blank: Fill a cuvette with the appropriate volume of DI

water instead of sample. Like for the sample cuvette accomplish exactly all

further working steps accordingly to the LCK working procedure.

Note: Only one reagent blank per each test charge is necessary. Store it darkly.

2.

Press POWER to switch on the instrument. COD LCK 014/114 with POCKET

COLORIMETER II 600 nm, COD LCK 314/614 with POCKET COLORIMETER

II 450 nm.

3.

Make sure that the correct measuring range is selected!

Otherwise press MENU. "SEL" is displayed. A flashing arrow indicates the

current measuring range (1 or 2). Press GREEN key to select measuring range.

Press MENU to get back to the measuring mode.

4.

Insert the reagent blank into the cell holder equipped with the cuvette adapter

1 cm.

Note: Make sure that the cuvette adapter is positioned correctly.

5.

Place the instrument cap with the HACH LANGE label over the cell holder to

cover the cuvette.

6.

Press BLUE key. "0" (LCK 014/114) or "0.0" (LCK 314/614) is displayed.

7.

Remove the instrument cap and insert the sample cuvette.

8.

Place the instrument cap over the cell holder to cover the cuvette.

9.

Press GREEN key.