Hydrogen sensor, Ozone sensor – Hach-Lange ORBISPHERE 31xxx Basic User Manual User Manual

Page 17

Advertising
background image

Hydrogen sensor

Problem

Probable cause

Possible solution

Sensor won't calibrate, even after
thorough servicing.

Repeated calibrations go beyond
"expected limits" of instrument.

MOCA 3600 only: Select membrane
from "Options/Membrane" menu.
Then, calibrate the sensor.

"0000" H

2

levels displayed.

Wrong reading scale "XXXX"
selected for display unit.

Change reading scale by selecting
"X.XXX, XX.XX or XXX.X".

Shorter than expected sensor
operation in relatively high H

2

concentration.

High H

2

concentrations require

more work from electrochemical
sensor.

Shut off analyzer when not needed.

Unexpected or inaccurate H

2

readings.

High residual current.

If concentration is significantly
higher than low limit, try a sensor
service.

Ozone sensor

When the O

3

sensor has been properly calibrated using the ORBISPHERE measuring instrument,

the sensor has to settle down for up to 24 hours when used in very low O

3

concentration conditions.

Problem

Probable cause

Possible solution

Sensor won't calibrate, even after
thorough servicing.

Repeated calibrations go beyond
"expected limits" of instrument.

MOCA 3600 only: Select membrane
from "Options/Membrane" menu.
Then, calibrate the sensor.

Instrument internal barometric
pressure sensor needs calibration.

Calibrate internal barometer against
a certified barometer. Do not correct
for sea level !

Wet membrane interface.

Wipe dry with a tissue and re-
calibrate.

"0000" O

3

levels displayed.

Wrong reading scale "XXXX"
selected for display unit.

Change reading scale by selecting
"X.XXX, XX.XX or XXX.X".

Unexpected or inaccurate dissolved
O

3

readings.

High residual current.

If concentration is significantly
higher than low limit, try a sensor
service.

Insufficient flow rate.

Regulate flow equivalent to
membrane specified levels.

Length of sample line allows O

3

time to react.

Reduce length of sample tubing.

Doesn't match lab samples.

Take samples at close proximity to
sensor.

English 17